T
      
        

 
cool. Retest.
or
 
       







           


      


         
         
        
          
electrode.



It is possible to desensitize the ORP sensor when measuring
high ORP solutions. To check the ORP sensor functioning, take
an ORP reading of Myron L pH/ORP sensor storage solution. If
the reading is outside the range of 350-400 mV, clean ONLY the
platinum ORP electrode with Myron L ORP Conditioner solution-
soaked cotton swab, being careful not to touch the swab to the
glass bulb of the pH sensor.