III. RULES of OPERATION A. OperationUsing the instrument is simple:
b. A press of displays resistivity with units on the right. On the left
UP or DOWN KeysWhile measuring in any parameter, the or keys activate While in CAL mode, the k
10
113. Press the or PP
stream of the sol
13 D. Procedure to Select a SolutionNOTE:
14 Disabling Temperature Compensation b. With “User” selected, press
VII. CALIBRATION A. Generally, calibration is recommended about once per month with
BUFFER°CRATIO% /ORP RES TDSpHCOND09nov10Reference Junction underGlass pH BulbThese Measurement keys will: • Turn instrument on • Measure par
16 Calibration LimitsThere are calibration limits. A nominal “FAC” value is an ideal value stored by the factory. Attempts to calib
e. Press calibration sequence for this particular solution type. 3.
4. Press to enter calibration mode. The “CAL”, “BUFFER” and “7 Displayed value will be the u
Figure 7 BUFFERpHCALFigure 8pHBUFFERCALNOTE:Acd” and “bAS” indicators are blinking, the unit is indicating an error and needs either an aci
VIII. CALIBRATION INTERVALSThere is no simple answer as to how often one should calibrate an instrument. The Ultrameter II is desi
D. 1. Keep the sensor wet with Myron L Storage Solution.
C. to clear that memory. This space will be the
3. Press to initiate. “CAL4. Press or to change the time.
or to change the day. to accept Figure 15CAL C.
Figure 18Figure 194. Press to switch units. to accept unit preference for all temperature readings.NOTE:
BUFFER°CRATIO% /ORP RES TDSpHCOND09nov10Reference Junction underGlass pH BulbThese Measurement keys will: • Turn instrument on • Measure par
XIII. CELL CHECK
to display the stored memory records.3. Press repeatedly until you pass the CELL ch location. The display will show “Auto oFF
B. Setting User mode Calibration “Linc”The Linc function sets or “links” the calibration gain factor of a Standard Solution to the User solution
measurement keys will exit without changing the setting. User mode “Linc” is now complete. The User mode will now use the calibration gain const
30XVI. bluDock™ WIRELESS DATA TRANSFER INSTRUCTIONSNOTE: Bluetooth®bluDock Bluetooth module is a reg
31NOTE: “PC Ini” may momentarily be Figure 32 4. Add bluDock to your Bluetooth devices
XVII. CARE and MAINTENANCEUltrameter IIs should be rinsed with clean water after use. Solvents should be avoided. Shock damage from a fall may caus
33beallowed to dry out.a
34XVIII. TROUBLESHOOTING CHARTSymptom Possible Cause Corrective ActionNo display, even though measurement key pressedBattery weak or not connected.
XVIII. TROUBLESHOOTING CHARTSymptom Possible Cause Corrective ActionNo display, even though measurement key pressedBattery weak or not connected.
36XIX. ACCESSORIESNOTE: MSDSs are available on the Myron L website for all solutions:http://www.myronl.com/main/Material_Safety_DS_DL.htm A.
E. Soft Protective Carry CasesPadded Nylon carrying case features a belt clip for hands-free mobility. Two colors to choose from:
Chart 1 0 5 10 15 20 25 30 35 40 45 50 55 601.500%1.600%1.700%1.800%1.900%2.000%2.100%2.200%2.300%2.400%2.500%KCl %
7%Chart 255(1)%(2)%0%1%2%3%4%5%6%0 5 10 15 202530 35 40 45 50TemperatureNaCl error with KCl tempco442 error with KCl tempco E. Other SolutionsA sa
40 A. Once the effect of temperature is removed, the compensated conductivity
41the User
There is a ratio of TDS to compensated conductivity for any solution, which varies with concentration. The ratio is set during calibratio
43still produce an offset voltage, but using the same materials to connect offset
44 c. Temperature Compensation
C. Free Chlorine 1.Chlorine is a germicide used in the disinfection of drinking water,
1I. INTRODUCTIONThank you for selecting the feature-packed Ultrameter II™, one of the Myron L Company’s latest in an increasing line of instruments
46XXV. GLOSSARYAnions Negatively charged ions. See Solution Characteristics, pg. 40.Algorithm A procedure for solving a mathematical problem.
• Accuracy of ±1% of READING ±.2% at Calibration Point• Reliable Repeatable Results• KCl, NaCl and 442™ Natural Water Modes• Automatic Temperat