
FEATURES and SPECIFICATIONS
A. Features
• Superior resolution 4 digit LCD displays full 9999 µS/ppm.
• Accuracy of ±1% of reading (not merely full scale).
• All electrodes are internal for maximum protection.
• Latest 4 electrode cell technology.
• Waterproof to 3 feet/1 meter.
• Autoranging conductivity/TDS/resistivity.
• Memory saves 20 readings.
• Factory calibrations stored in microprocessor.
• 3 conductivity/TDS solution conversions preprogrammed into
microprocessor.
• USER feature allows:
Programming your own cond/TDS conversion factor.
Programming your own temperature compensation factor.
Disabling temperature compensation.
B. General Specifications
Display 4 Digit LCD
Dimensions (LxWxH) 7.7x2.7x2.5 in.
196x68x64 mm
Weight 12oz./341g
Case Material VALOX*
Cond/Res/TDS Cell Material VALOX*
Cond/Res/TDS Cell Capacity 0.2 oz./5 ml
Power 9V Alkaline Battery
Battery Life >100 Hours/5000 Readings
Operating/Storage Temperature 32-132°F/0-55°C
Protection Ratings IP67/NEMA 6 (waterproof to
3 feet/1 meter)
* ™ GE Corp.
Additional information available on our web site at:
www.myronl.com
2
C. Specification Chart
D. Warranty/Service
If, in the opinion of the factory, failure was due to materials or
workmanship, repair or replacement will be made without charge. A
reasonable service charge will be made for diagnosis or repairs due to
normal wear, abuse or tampering. This warranty is limited to the repair or
replacement of the Ultrameter only. The Myron L Company assumes
other responsibility or liability.
E. Ultrameter Models
ULTRAMETER MODELS 3P 4P 6P
PARAMETERS pH/ORP/Temp. Conductivity/TDS Conductivity/TDS/pH
Resistivity/Temp. Resistivity/ORP/Temp.
3
0-71°C
32 - 160°F
0-71°C
32 - 160°F
0-71°C
32 - 160°F
MYRON L COMPANY
2450 Impala Drive
Carlsbad, CA 92010
USA
760-438-2021
±1 % of reading**
0-71°C
32 - 160 °F
Ω
Ω
0-9999 µS
10-200 mS
in 5 autoranges
±1 % of reading* ±1 % of reading*
0-9999 ppm
10-200 ppt
in 5 autoranges
0.01 (<100 ppm)
0.1 (<1000 ppm)
1.0 (>1000 ppm)
Auto
Temperature
Compensation
Adjustable
Cond/TDS Ratio
Factor
0.20 - 7.99
Cond/TDS
Ratios
Preprogrammed
KCl, NaCl, 442™
0 - 9.99%/°CAdjustable
Temperature
Compensation
0 - 9.99%/°C 0 - 9.99%/°C
0.01 (<100 µS)
0.1 (<1000 µS)
1.0 (>1000 µS)
0.01 (<100K ohmsΩ
)
0.1 (<1000K ohmsΩ
)
1.0 (>1 MΩ)
0.1 °C/FResolution
Ranges
±0.1 °CAccuracy
Conductivity TDS Resistivity
10K - 30M ohms
Temperature
4P
* up to 100mS/ppt; 100 - 200mS/ppt: ± 1 - 2% typ.
**10K ohms - 10M ohms
Ω Ω
All Myron L Ultrameters have a 2 year warranty except for pH sensors
which have a 6-month limited warranty. If an instrument fails to operate
properly, see the Troublshooting Chart, pg. 21. The battery and pH/ORP
sensor are user replaceable. For other service, return the instrument
prepaid to the Myron L Company.
Comentarios a estos manuales