  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 18
User) is already the type desired. If not:
1. Press , or to select the parameter on which
you wish to change the solution type.
2. Press and hold key
Figure 1
KCl
442
NaCl
User
for 3 seconds to make “SEL
appear (see Figure 1). For
demonstration purposes, all
4 solution types are shown
simultaneously.
3. Use the or key to select type of solution desired
(ref. Solution Characteristics, pg. 42). The selected solution
type will be displayed: KCl, NaCl, 442 or User.
4. Press to accept new solution type.
E. Application of User Solution Type
1. User Programmable Temperature Compensation
(Tempco)
This feature allows you to change your Ultrameter II’s temperature
compensating factor to another factor between 0-9.99%/°C (ref.
Temperature Compensation, pg. 39). This feature does not apply to pH
or ORP.
a. As in D. Procedure to Select a Solution above, select “User” mode.
b. With User mode now selected, press . You may now
adjust a temperature compensation from .00%/°C to 9.99%/°C,
by pressing or .
See example in Figure 2.
c. Press twice to skip
Figure 2
°C% /
User
COND
calibration adjustment and
accept the new tempco (3
times if in TDS mode). You
are now ready to measure samples with your new temperature
compensation factor.
2. Disabling Temperature Compensation
a. Select User mode (ref. Procedure to Select a Solution, pg. 14).
15