2. Press to display the stored memory records.
3. Press repeatedly until you pass the CELL ch location.
The display will show “Auto oFF” (Figure 24).
4. Press
to initiate. CAL
Figure 25
CAL
will be displayed along with
15 SEC” or current Auto
oFF value (see Figure 25).
5. Press or to
change the amount of time
Figure 26
CAL
(see Figure 26). Maximum
time of 75 seconds is shown.
6. Press to accept the
change (new time).
XV. USER MODE CALIBRATION LINC™ FUNCTION
The Linc™ function allows easy calibration when in User mode and the
user does not have a user standard solution to calibrate the instrument.
This function will ensure more repeatable and accurate measurements
than many other calibration methods. It is recommended that this
function be used to provide the highest degree of condence when the
Ultrameter II is used in User mode. When Linc is used, the User mode
is linked to another standard, i.e., if User and KCl are linked, a KCI
standard solution is used to calibrate the instrument. It is that simple.
A. Calibration of Ultrameter II for use in User mode
1. Press
or key.
2. Calibrate the unit using a Standard Solution (ref.
CALIBRATION, pg. 17).
3. Place the Ultrameter II in User mode (ref. SOLUTION
SELECTION, pg. 14).
4. Verify/Set the calibration linc. (See B. Setting User
Mode Calibration “Linc”, pg. 30.)
29